Journal Article


A high sensitivity and low power circuit for the measurement of abnormal blood cell levels

Abstract

This paper describes a technique to detect blood cell levels based on the time-period modulation of a relaxation oscillator loaded with an Inter Digitated Capacitor (IDC). A digital readout circuit has been proposed to measure the time-period difference between the two oscillators loaded with samples of healthy and (potentially) unhealthy blood. A prototype circuit was designed in 65nm CMOS technology and post-layout simulations shows 15.25aF sensitivity. The total circuit occupies 2184µm2 silicon area and consumes 216µA from a 1V power supply.

Attached files

Authors

Nagulapalli, Rajasekhar
Hayatleh, Khaled
Barker, Steve
Tammam, Abdallah
Lidgey, Francis J.
Yassine, Nabil

Oxford Brookes departments

Faculty of Technology, Design and Environment\School of Engineering, Computing and Mathematics

Dates

Year of publication: 2019
Date of RADAR deposit: 2019-05-07


Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License


Related resources

This RADAR resource is the Accepted Manuscript of A high sensitivity and low power circuit for the measurement of abnormal blood cell levels

Details

  • Owner: Joseph Ripp
  • Collection: Outputs
  • Version: 1 (show all)
  • Status: Live
  • Views (since Sept 2022): 340