Journal Article


Analysis and design of a high precision- high output impedance tissue current driver for medical applications

Abstract

This paper describes the design and operation of a high output impedance tissue current driver circuit, for use in medical electronics, such as Electrical Impedance Tomography (EIT). This novel architecture was designed for implementation in bipolar technology, to meet the specifications for EIT, namely operating frequency range 10 kHz–1 MHz with a target output resistance of 16 MW. Simulation results are presented, showing that the current source more than met the minimum specification for EIT.

Attached files

Authors

Terzopoulos, N
Hayatleh, K
Sebu, C
Lidgey, FJ
Ben-Esmael, M
Tammam, A
Barker, Steve

Oxford Brookes departments

Faculty of Technology, Design and Environment\Department of Mechanical Engineering and Mathematical Sciences

Dates

Year of publication: 2014
Date of RADAR deposit: 2016-02-17



Related resources

This RADAR resource is the Accepted Manuscript of Analysis and design of a high precision- high output impedance tissue current driver for medical applications

Details

  • Owner: Rosa Teira Paz
  • Collection: Outputs
  • Version: 1 (show all)
  • Status: Live
  • Views (since Sept 2022): 524